ГлавнаяУязвимости → CVE-2024-48970
9.3критическая

CVE-2024-48970

Описание

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

CVSS
Балл9.3 — критическая
ВекторCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Источники
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01